Enter at least 2 characters to start the search.

Digital combined pH/ORP sensor
Memosens CPS76D

Memosens CPS76D - Combined pH/ORP glass sensor for demanding processes and hygienic applications ©Endress+Hauser
Memosens CPS76D - For simultaneous pH and ORP measurement ©Endress+Hauser
Memosens CPS76D - Detail view platinum element ©Endress+Hauser

The pH-value of a liquid can be calculated using the potentiometric measurement principle. This video shows what it is about and how this measuring principle works.

Memosens converts the measured value to a digital signal and transfers it to the transmitter without a contacting connection. This video shows how Memosens optimizes process control and maintenance.

Memosens glass electrode for the chemical and life sciences industries

Memosens CPS76D is the expert for simultaneous pH and ORP measurement in demanding processes and hygienic applications, offering a better process control and increased quality. The sensor measures reliably at high temperatures and is non-toxic assuring highest safety of your products. Thanks to Memosens digital technology, CPS76D combines maximum process and data integrity with simple operation. It resists corrosion and moisture, enables lab calibration and facilitates predictive maintenance.

Quick selection

Choose from our suggested options
  • <%label%>
  • <%feature%>
  • Process Connection
  • <%message%>

or

Individual configuration

Continue your selection from all possible options
  • Benefits

    • Simultaneous measurement of pH, ORP and rH values (in rH mode) for a better process overview

    • Additional platinum electrode enables constant monitoring of reference impedance and thus sensor quality

    • Robust and resistant to poisoning thanks to new...

  • Field of application

    • Process technology and monitoring of processes with:
      - Quickly changing pH values
      - High proportion of electrode poisons such as H2S

    • Chemical processes

    • Hygienic and sterile applications:
      - Fermenters
      - Biotechnology
      - Life sciences

    With ATEX, FM, and CSA...

Features and specifications

Back

Details

  • Measuring principle

    Glass electrode

  • Application

    Hygienic and sterile applications, process

  • Characteristic

    CIP/SIP and autoclavable - high accuracy - biocompatibility of the gel inside - version upside-down installation available - pressurized reference (6 bar)
    Pt-disc as additional ORP element.
    rH measurment and control of the reference impedance.

  • Measurement range

    pH 0-14

  • Measuring principle

    - ceramic diaphragm - Ion trap

  • Design

    - All shaft lengths with temperature sensor - advanced gel technology - digital electrode with Memosens technology

  • Material

    glass / ceramic

  • Dimension

    Diameter: 12 mm (0.46 inch)
    Shaft lengths: 120, 225, 360 and 425 mm
    (4.68, 8.77, 14.04 and 16,57 inch)

  • Process temperature

    -15°C ... 140°C
    (5°F - 284°F)

  • Process pressure

    max. 13 bar
    (188 psi)depends on selected version

  • Temperature sensor

    NTC 30k

  • Connection

    Inductive, digital connection head with Memosens technology.

  • Ingres protection

    IP68

Documents / Manuals / Software

Accessories / Spare parts

Nr
Order code
Specification