This product is no longer available.
Please check new generation in the section predecessor / successor or contact your local Endress+Hauser entity.
-
Successor
-
Standard products
- Reliable, robust and low-maintenance
Technical excellence
Simplicity
-
Specialized products
- Designed for demanding applications
Technical excellence
Simplicity
Variable
FLEX selections Technical excellence Simplicity - F
- L
- E
- X
Fundamental selection
Meet your basic measurement needs
Technical excellence Simplicity - F
- L
- E
- X
Lean selection
Handle your core processes easily
Technical excellence Simplicity - F
- L
- E
- X
Extended selection
Optimize your processes with innovative technologies
Technical excellence Simplicity - F
- L
- E
- X
Xpert selection
Master your most challenging applications
Technical excellence Simplicity Variable
-
-
Measuring principle
Coriolis
-
Product headline
The easy‐to‐clean flowmeter with self‐drainable single‐tube system and a compact or remote transmitter. Dedicated for applications requiring optimal cleanability under hygienic conditions.
-
Sensor features
Reduced installation costs – fully self‐drainable tube design enables compact horizontal mounting. Fewer process measuring points – multivariable measurement (flow, density, temperature). Spacesaving installation – no in/outlet run needs. Large range of hygienic process connections. 3-A conform.
-
Transmitter features
Cost‐effective – dedicated design for standard applications. Safe operation – display provides easy readable process information. Fully industry compliant – IEC/EN/NAMUR. 2‐line backlit display with push buttons. Device in compact or remote version.
-
Nominal diameter range
DN 8 to 50 (⅜ to 2")
-
Wetted materials
Measuring tube: 1.4539 (904L)
Connection: 1.4435 (316L); 1.4404 (316/316L) -
Measured variables
Mass flow, density, temperature, volume flow, corrected volume flow, reference density
-
Max. measurement error
Mass flow (liquid): ±0.15 %
Volume flow (liquid): ±0.15 %
Mass flow (gas): ±0.5 %
Density (liquid): ±0.0005 g/cm³ -
Measuring range
0 to 70 000 kg/h (0 to 2570 lb/min)
-
Max. process pressure
PN 63, Class 300, 40K
-
Medium temperature range
–50 to +150 °C (–58 to +302 °F)
-
Ambient temperature range
Standard: –20 to +60 °C (–4 to +140 °F)
Option: –40 to +60 °C (–40 to +140 °F) -
Sensor housing material
1.4301 (304), corrosion resistant
-
Transmitter housing material
Powder‐coated die‐cast aluminium
1.4301 (304), sheet
CF3M (316L), cast -
Degree of protection
IP67, type 4X enclosure. Remote transmitter: IP67, type 4X enclosure
-
Display/Operation
2‐line backlit display with push buttons
Configuration via local display and operating tools possible -
Outputs
3 outputs:
0‐20 mA (active)/4‐20 mA (active/passive)
Pulse/frequency/switch output (passive) -
Inputs
Status input
-
Digital communication
HART
PROFIBUS PA -
Power supply
DC 16 to 62 V
AC 85 to 260 V (45 to 65 Hz)
AC 20 to 55 V (45 to 65 Hz) -
Hazardous area approvals
ATEX, IECEx, FM, CSA, NEPSI
-
Product safety
CE, C-tick, EAC marking
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR
-
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
-
Hygienic approvals and certificates
3-A, FDA
-
Measuring principle
Coriolis
-
Product headline
The easy‐to‐clean flowmeter with self‐drainable single‐tube system and a compact or remote transmitter. Dedicated for applications requiring optimal cleanability under hygienic conditions.
-
Sensor features
Reduced installation costs – fully self‐drainable tube design enables compact horizontal mounting. Fewer process measuring points – multivariable measurement (flow, density, temperature). Spacesaving installation – no in/outlet run needs. Large range of hygienic process connections. 3-A conform.
-
Transmitter features
Cost‐effective – dedicated design for standard applications. Safe operation – display provides easy readable process information. Fully industry compliant – IEC/EN/NAMUR. 2‐line backlit display with push buttons. Device in compact or remote version.
-
Nominal diameter range
DN 8 to 50 (⅜ to 2")
-
Wetted materials
Measuring tube: 1.4539 (904L)
Connection: 1.4435 (316L); 1.4404 (316/316L) -
Measured variables
Mass flow, density, temperature, volume flow, corrected volume flow, reference density
-
Max. measurement error
Mass flow (liquid): ±0.15 %
Volume flow (liquid): ±0.15 %
Mass flow (gas): ±0.5 %
Density (liquid): ±0.0005 g/cm³ -
Measuring range
0 to 70 000 kg/h (0 to 2570 lb/min)
-
Max. process pressure
PN 63, Class 300, 40K
-
Medium temperature range
–50 to +150 °C (–58 to +302 °F)
-
Ambient temperature range
Standard: –20 to +60 °C (–4 to +140 °F)
Option: –40 to +60 °C (–40 to +140 °F) -
Sensor housing material
1.4301 (304), corrosion resistant
-
Transmitter housing material
Powder‐coated die‐cast aluminium
1.4301 (304), sheet
CF3M (316L), cast -
Degree of protection
IP67, type 4X enclosure. Remote transmitter: IP67, type 4X enclosure
-
Display/Operation
2‐line backlit display with push buttons
Configuration via local display and operating tools possible -
Outputs
3 outputs:
0‐20 mA (active)/4‐20 mA (active/passive)
Pulse/frequency/switch output (passive) -
Inputs
Status input
-
Digital communication
HART
PROFIBUS PA -
Power supply
DC 16 to 62 V
AC 85 to 260 V (45 to 65 Hz)
AC 20 to 55 V (45 to 65 Hz) -
Hazardous area approvals
ATEX, IECEx, FM, CSA, NEPSI
-
Other approvals and certificates
3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR, SIL
PED, CRN
3-A, FDA -
Product safety
CE, C-tick, EAC marking
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR
-
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
-
Hygienic approvals and certificates
3-A, FDA
-
Measuring principle
Coriolis
-
Product headline
The easy‐to‐clean flowmeter with self‐drainable single‐tube system and a compact or remote transmitter. Dedicated for applications requiring optimal cleanability under hygienic conditions.
-
Sensor features
Reduced installation costs – fully self‐drainable tube design enables compact horizontal mounting. Fewer process measuring points – multivariable measurement (flow, density, temperature). Spacesaving installation – no in/outlet run needs. Large range of hygienic process connections. 3-A conform.
-
Transmitter features
Cost‐effective – dedicated design for standard applications. Safe operation – display provides easy readable process information. Fully industry compliant – IEC/EN/NAMUR. 2‐line backlit display with push buttons. Device in compact or remote version.
-
Nominal diameter range
DN 8 to 50 (⅜ to 2")
-
Wetted materials
Measuring tube: 1.4539 (904L)
Connection: 1.4435 (316L); 1.4404 (316/316L) -
Measured variables
Mass flow, density, temperature, volume flow, corrected volume flow, reference density
-
Max. measurement error
Mass flow (liquid): ±0.15 %
Volume flow (liquid): ±0.15 %
Mass flow (gas): ±0.5 %
Density (liquid): ±0.0005 g/cm³ -
Measuring range
0 to 70 000 kg/h (0 to 2570 lb/min)
-
Max. process pressure
PN 63, Class 300, 40K
-
Medium temperature range
–50 to +150 °C (–58 to +302 °F)
-
Ambient temperature range
Standard: –20 to +60 °C (–4 to +140 °F)
Option: –40 to +60 °C (–40 to +140 °F) -
Sensor housing material
1.4301 (304), corrosion resistant
-
Transmitter housing material
Powder‐coated die‐cast aluminium
1.4301 (304), sheet
CF3M (316L), cast -
Degree of protection
IP67, type 4X enclosure. Remote transmitter: IP67, type 4X enclosure
-
Display/Operation
2‐line backlit display with push buttons
Configuration via local display and operating tools possible -
Outputs
3 outputs:
0‐20 mA (active)/4‐20 mA (active/passive)
Pulse/frequency/switch output (passive) -
Inputs
Status input
-
Digital communication
HART
PROFIBUS PA -
Power supply
DC 16 to 62 V
AC 85 to 260 V (45 to 65 Hz)
AC 20 to 55 V (45 to 65 Hz) -
Hazardous area approvals
ATEX, IECEx, FM, CSA, NEPSI
-
Other approvals and certificates
3.1 material, calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR, SIL
PED, CRN
3-A, FDA -
Product safety
CE, C-tick, EAC marking
-
Functional safety
Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511
-
Metrological approvals and certificates
Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025), NAMUR
-
Pressure approvals and certificates
PED, CRN
-
Material certificates
3.1 material
-
Hygienic approvals and certificates
3-A, FDA
-
Measuring principle
Coriolis
-
Characteristic / Application
Coriolis mass and density flowmeter for the food and beverage industry.
-
Ambient temperature
-20...+65°C
(-4...140°F) -
Process temperature
-50...+150°C
(-58...+302°F) -
Process pressure absolute
PN 16...40
Cl 150...300
JIS 20...40K -
Wetted parts
904L/1.4539
-
Output
4...20mA
Pulse/Frequency
Status -
Certificates / Approvals
ATEX
FM
CSA IECEx TIIS NEPSI